Pace 300 EXTERNAL PACEMAKERS
Pace 300 EXTERNAL PACEMAKERS
Pace 300 EXTERNAL PACEMAKERS
Triple Chamber External Pacemaker
Oscor’s PACE 300 is the world’s first temporary three-chamber cardiac pacemaker featuring bi-ventricular pacing capability with individually programmable ventricular pacing and sensing channels and programmable inter-ventricular stimulation delay. It is designed for cardiac resynchronization therapy with one atrial and two independent ventricular channels. The available trigger modes and the adjustable interventricular stimulation delay provide a resynchronization option of both ventricles for different heart diseases: Arrhythmia and cardiac conduction disorders, Bradycardia after cardiac surgery, Pacemaker implant or replacement procedures.
FEATURES
Easy setup of stimulation parameters by ergonomically designed dials and soft keys
High rate function for rapid atrial pacing of up to 1,000 ppm (adjustable while applying rapid atrial pacing therapy)
Activate emergency stimulation mode (D00, 12 V, 80 ppm, BV) by pressing “EMERGENCY” button
Convenient measurement of P/R wave peak amplitudes while pressing “PAUSE” button
PVARP and MTR depending on selected stimulation rate
Optional AUTO SENSE function which automatically tracks P/R wave peak amplitudes and adjusts atrial and ventricular sensitivities accordingly
Automatic mode switch to asynchronous modes upon detection of interference
Continuous surveillance of battery exhaustion and pacing lead system integrity
Large backlit display informs quickly about battery status, pacing mode, parameter settings, activated automatic functions and lead system safety surveillance
LEDs indicate whether knobs for ventricular output and sensitivity apply to RV + LV, RV only or LV only
Customize, save and recall pre-set parameter values for quick setup
Multilingual menu options, including English, Spanish, German, French, Italian or Swedish
Standard 9 V Alkaline battery provides at least 8 days of continuous operation (DDD, standard parameters)
Optional AC power supply saves battery power. Battery must remain in battery compartment in case of AC power shortage
Patient cable system featuring patented connections with protected pins for convenient, reliable electrical connections to increase patient safety
Standard 2 year limited warranty
TECHNICAL SPECIFICATIONS:
Set includes:
Standard 9V battery, 3 ATAR extension cables, and a carrying case.
Stimulation Rate 30 ... 220 ppm
Maximum Tracking Rate 80 ... 230 ppm
Atrial overdrive stimulation 70 ... 1000 ppm
Stimulation amplitude 0.1 ... 18 V ± 10%
Pulse duration 0.05 ... 1.5 ms ± 5%
Sensitivity Atrium: 0.2 ... 20 mV ± 10%, infinite, Ventricle: 1.0 ... 20 mV ± 10%, infinite
Refractory Period 250 ... 400 ms 100 ... 500 ms 250 ms
A-V delay 5 ms ... 400 ms ± 5% ± 2 ms
V-V Stimulation Delay VVD 5 ms ... 80 ms ± 5% ± 2 ms
Pacing modes DDD (BV), DDD (LV), DDD (RV), VVI (BV), VVV (LV), VVI (RV), AAI, VDD (BV), VDD (RV), VDD (LV), DDT, DDI, DVI, DAI, DDD +AT, DAT, DOO, ODO, VVT, VOO, AAT, AOO, VDI, VAT